MOLECULAR BIOLOGYCell BiologyBiology Calculator
🧬

Protein Concentration from A280

BCA/Bradford assay or A280: C = A/(ε×l). mg/mL. Extinction coefficient from sequence or database.

Concept Fundamentals
C = A/(ε×l)
A280
ε ≈ 43,824
BSA
ε ≈ 210,000
IgG
Common unit
mg/mL
Calculate Protein ConcentrationBCA/Bradford and A280 methods

Why This Biology Metric Matters

Why: Accurate protein concentration is essential for enzyme assays, Western blots, and structural biology.

How: A280: C (mg/mL) = A280/(ε×l). ε from ProtParam or known values. BCA/Bradford: standard curve from known protein.

  • A280 assumes Trp/Tyr content. BSA ε ≈ 43,824 M⁻¹cm⁻¹.
  • BCA and Bradford more sensitive; A280 for pure proteins.
  • Buffer absorbance at 280 nm can interfere; blank carefully.

Sample Scenarios

BSA Standard Measurement

Typical BSA concentration measurement at 280nm

Click to load

IgG Purification Check

Checking IgG concentration after protein purification

Click to load

Lysozyme Activity Assay

Measuring lysozyme concentration for enzyme activity

Click to load

Custom Protein Analysis

Custom protein with known extinction coefficient

Click to load

Tryptophan-Rich Protein

Protein with high tryptophan content (high ε)

Click to load

Low Concentration Sample

Diluted sample requiring careful measurement

Click to load

Enter Measurement Parameters

🔬 Protein Selection

Select protein type or choose custom

📊 Measurement Parameters

Measured absorbance value (typically 0.1-1.0)
Cuvette pathlength (usually 1.0 cm)
Dilution factor (1 = no dilution, 10 = 10x diluted)

💧 Optional: Sample Volume

Total sample volume to calculate total protein amount

For educational use only. Always confirm dosages and care with a licensed veterinarian.

🧬 Biology Facts

📐

C = A280/(ε×l). Beer-Lambert for proteins. ε in M⁻¹cm⁻¹.

— A280

🧪

BCA and Bradford: colorimetric, standard curve. More sensitive.

— Assays

🔬

BSA ε ≈ 43,824; IgG ≈ 210,000. ProtParam for custom.

— Extinction

📊

mg/mL common. Convert to molar: MW from sequence.

— Units

What is Protein Concentration Measurement?

Protein concentration measurement is a fundamental technique in biochemistry and molecular biology. The UV absorbance method at 280nm is one of the most common, rapid, and non-destructive methods for determining protein concentration. It relies on the fact that proteins contain aromatic amino acids (tryptophan, tyrosine, and phenylalanine) that absorb ultraviolet light at 280nm.

🔬

UV Absorbance Method

Measures absorbance at 280nm where aromatic amino acids absorb light. Fast, non-destructive, and requires minimal sample.

📊

Beer-Lambert Law

The fundamental principle: absorbance is proportional to concentration, pathlength, and extinction coefficient.

🎯

Extinction Coefficient

Unique to each protein, depends on the number and type of aromatic amino acids. Higher ε = more sensitive detection.

How the Beer-Lambert Law Works

The Beer-Lambert Law describes the relationship between light absorption and the properties of the material through which light passes:

A = ε × b × C

Where:

A = Absorbance (unitless, typically 0-2)

ε = Molar extinction coefficient (M⁻¹ cm⁻¹)

b = Pathlength (cm, usually 1.0 cm for standard cuvettes)

C = Molar concentration (M)

To calculate protein concentration in mg/mL, we rearrange and multiply by molecular weight and dilution factor:

C = (A / (ε × b)) × MW × n

Where:

C = Mass concentration (mg/mL)

MW = Molecular weight (g/mol)

n = Dilution factor (if sample was diluted)

Key Considerations

  • Keep absorbance 0.1–1.0 for accurate measurements
  • Always use buffer blank to subtract background
  • Nucleic acids absorb at 280nm—check A260/A280 ratio (~0.6 for pure protein)
  • Verify cuvette pathlength (usually 1.0 cm)

When to Use UV Absorbance vs Other Assays

✅ UV Absorbance (280nm) - Best For:

  • • Pure protein samples (no nucleic acids)
  • • Known protein with known extinction coefficient
  • • Quick, non-destructive measurements
  • • Monitoring protein purification
  • • When sample volume is limited
  • • Real-time concentration monitoring

⚠️ Consider Other Methods When:

  • • Sample contains nucleic acids (use Bradford, BCA, or Lowry)
  • • Protein has very low/no aromatic amino acids
  • • Unknown protein without known ε
  • • Need absolute accuracy (use amino acid analysis)
  • • Sample has interfering compounds
  • • Very low concentrations (<0.1 mg/mL)

Common Protein Extinction Coefficients

Extinction coefficients vary significantly based on aromatic amino acid content. Tryptophan contributes most (~5500 M⁻¹ cm⁻¹), followed by tyrosine (~1490 M⁻¹ cm⁻¹), and phenylalanine (~200 M⁻¹ cm⁻¹).

ProteinExtinction Coefficient (ε)Molecular WeightNotes
BSA (Bovine Serum Albumin)43,824 M⁻¹ cm⁻¹66,463 g/molCommon standard protein, widely used in biochemical assays
IgG (Immunoglobulin G)210,000 M⁻¹ cm⁻¹150,000 g/molAntibody protein, commonly purified from serum
Lysozyme37,901 M⁻¹ cm⁻¹14,000 g/molSmall antimicrobial enzyme, high extinction coefficient
Insulin6,335 M⁻¹ cm⁻¹5,734 g/molHormone protein, low extinction coefficient
Streptavidin176,000 M⁻¹ cm⁻¹55,000 g/molBiotin-binding protein, very high extinction coefficient

Troubleshooting Common Issues

❌ Contamination Issues

Problem: Nucleic acids (DNA/RNA) also absorb at 280nm, causing overestimation.
Solution: Use A260/A280 ratio. Pure protein should have ratio ~0.6. If ratio >0.8, consider Bradford or BCA assay.

⚠️ Interference from Buffers

Problem: Some buffers (Tris, imidazole, DTT) absorb at 280nm.
Solution: Use buffer blank or measure at different wavelength (205nm for peptide bonds).

📏 Pathlength Accuracy

Problem: Using wrong pathlength (e.g., 0.5 cm cuvette but entering 1.0 cm).
Solution: Always verify cuvette pathlength. Many modern cuvettes are 1.0 cm, but some are 0.5 cm or 1.0 mm.

✅ Optimal Absorbance Range

Best Practice: Keep absorbance between 0.1-1.0 for accurate measurements.
Too Low (<0.1): High noise, inaccurate. Dilute less or use longer pathlength.
Too High (>2.0): Non-linear response. Dilute sample further.

Frequently Asked Questions

Q: Why measure at 280nm specifically?

A: Proteins contain three aromatic amino acids that absorb UV light: tryptophan (peak ~280nm), tyrosine (peak ~275nm), and phenylalanine (peak ~258nm). The 280nm wavelength is chosen because it's where tryptophan and tyrosine absorb most strongly, providing the best sensitivity for most proteins. Tryptophan has the highest extinction coefficient (~5500 M⁻¹ cm⁻¹ per residue), making it the primary contributor to protein absorbance.

Q: How do I find the extinction coefficient for my protein?

A: Extinction coefficients can be found in several ways: (1) Literature values - many common proteins have published ε values, (2) Calculate from sequence - tools like ProtParam (ExPASy) calculate ε from amino acid composition, (3) Experimental determination - measure absorbance of known concentration, (4) Use Bradford/BCA assay first to establish concentration, then calculate ε. For proteins with unknown ε, the Bradford or BCA assay is recommended instead of UV absorbance.

Q: What if my absorbance is too high or too low?

A: Optimal absorbance range is 0.1-1.0. If absorbance >1.0: Dilute your sample (e.g., 1:10 or 1:100) and multiply the calculated concentration by the dilution factor. If absorbance <0.1: Use a longer pathlength cuvette (e.g., 1 cm instead of 0.5 cm), or concentrate your sample. Very low absorbance values have high noise and poor accuracy. Always use a buffer blank to subtract background absorbance.

Q: Can I use this method if my sample contains DNA or RNA?

A: Nucleic acids also absorb strongly at 260nm and weakly at 280nm, which can interfere with protein measurements. Check the A260/A280 ratio: Pure protein should be ~0.6, while nucleic acid contamination gives ratios >0.8. If contaminated, use alternative methods like Bradford assay, BCA assay, or Lowry assay. Alternatively, measure at 205nm (peptide bond absorption) which is less affected by nucleic acids.

Q: How accurate is the UV absorbance method?

A: Accuracy depends on several factors: (1) Known extinction coefficient (±5-10% if from literature), (2) Absorbance measurement precision (±2-5% with good spectrophotometer), (3) Pathlength accuracy (±1% for standard cuvettes), (4) Sample purity (nucleic acid contamination can cause 10-50% overestimation). Overall, expect ±10-15% accuracy for pure proteins with known ε. For highest accuracy, use amino acid analysis or quantitative mass spectrometry.

Q: What's the difference between mg/mL and molar concentration?

A: mg/mL (mass concentration) tells you how many milligrams of protein per milliliter of solution - useful for practical applications like loading gels or preparing solutions. Molar concentration (M or mM) tells you how many moles of protein molecules per liter - important for stoichiometric calculations, enzyme kinetics, and binding studies. To convert: Molar = (mg/mL) / (MW in g/mol) × 1000. The calculator provides both units for convenience.

Q: Can I measure protein concentration in crude cell lysates?

A: UV absorbance at 280nm is generally NOT recommended for crude lysates due to interference from nucleic acids, small molecules, and other cellular components. Use colorimetric assays instead: Bradford assay (fast, sensitive, but affected by detergents), BCA assay (more compatible with detergents, very sensitive), or Lowry assay (classic method, more time-consuming). After purification, UV absorbance becomes more reliable.

Q: How do I account for buffer absorbance?

A: Always use a buffer blank! Measure your buffer alone at 280nm and subtract this value from your sample absorbance. Some buffers absorb significantly: Tris-HCl (minimal), HEPES (minimal), imidazole (absorbs at 280nm - use 290nm instead), DTT/TCEP (absorb at 280nm - use reducing agent-free buffer for blank). For best results, use the same buffer composition for both sample and blank.

Comparison with Other Protein Assay Methods

MethodPrincipleSensitivityAdvantagesLimitations
UV Absorbance (280nm)Aromatic amino acid absorption0.1-1 mg/mLFast, non-destructive, no reagents neededRequires known ε, affected by nucleic acids
Bradford AssayCoomassie Blue dye binding0.01-1.4 mg/mLVery sensitive, fast, simple protocolAffected by detergents, variable between proteins
BCA AssayBiuret reaction + BCA enhancement0.0005-2 mg/mLMost sensitive, compatible with detergentsMore expensive, longer incubation time
Lowry AssayBiuret + Folin-Ciocalteu reaction0.01-1 mg/mLClassic method, well-establishedTime-consuming, many interfering substances
Amino Acid AnalysisComplete hydrolysis + HPLCAny concentrationMost accurate, absolute quantificationExpensive, time-consuming, requires expertise

Aromatic Amino Acids and Their Contribution

The UV absorbance of proteins at 280nm comes from three aromatic amino acids. Understanding their individual contributions helps explain why extinction coefficients vary so much between proteins:

Tryptophan (Trp, W)

Extinction coefficient: ~5,500 M⁻¹ cm⁻¹ per residue
Peak wavelength: 280nm
Contribution: Highest - primary contributor to most protein absorbance

Most proteins contain 1-2% tryptophan. Proteins rich in Trp (like antibodies) have very high ε values.

Tyrosine (Tyr, Y)

Extinction coefficient: ~1,490 M⁻¹ cm⁻¹ per residue
Peak wavelength: 275nm
Contribution: Moderate - significant contributor

More common than tryptophan (~3-4% of residues). Can be ionized at high pH, affecting absorbance.

Phenylalanine (Phe, F)

Extinction coefficient: ~200 M⁻¹ cm⁻¹ per residue
Peak wavelength: 258nm
Contribution: Low - minimal contribution at 280nm

Common amino acid (~4% of residues) but contributes little to 280nm absorbance due to low ε and different peak.

Calculating Extinction Coefficient: For a protein with known sequence, ε can be calculated as: ε = (nTrp × 5,500) + (nTyr × 1,490) + (nPhe × 200) M⁻¹ cm⁻¹, where n is the number of each residue. Disulfide bonds can also contribute slightly (~200 M⁻¹ cm⁻¹ per bond).

👈 START HERE
⬅️Jump in and explore the concept!
AI

Related Calculators